![Cytiva AKTA Pure](https://primary.jwwb.nl/public/p/r/t/temp-clapbsszbkpnwzseqljn/20241120_184137-2-standard.jpg?enable-io=true&enable=upscale&crop=799%2C760%2Cx0%2Cy155%2Csafe&width=61&height=58)
What is Siphoning and Why is it Important?
Siphoning occurs in a chromatography system due to the tubing flow path's configuration. It happens when the outlet point is positioned lower than the reservoir, creating a vertical distance that drives flow under gravitational pressure. In chromatography systems, another scenario can occur where one pump is active while the other is idle. This can lead to a phenomenon where the active pump creates a vacuum or low-pressure area that causes fluid to be drawn from the inlet line of the idle pump.
To counteract siphoning:
- Pump flow restrictor valves are installed.
- These valves remain closed under gravitational pressure and only open when the pump creates enough pressure to overcome it.
Importance:
- Prevents flow when pumps are idle.
- Avoids unwanted contamination between the A & B inlet lines.
When to Perform a Siphoning Test
- Annually: It is best practice to perform the test at least once a year.
- When Suspected Issues Arise: Indications of a faulty pump flow restrictor valve include:
- Increased intake from the reservoir.
- Inconsistent purification results.
- Unstable UV & Conductivity readings.
How to Perform the Siphoning Test
Performing the test is quick and straightforward, typically taking about 5 minutes. In addition to our step-by-step guide, we have created a short easy to follow video for your assistance.
Materials Required:
- A container with DI water.
- A suitably sized syringe to purge the system pumps.
1. Prepare the Setup:
- Place a container filled with DI water on top of the instrument and Immerse inlets A1 and B1 into the container.
![Cytiva AKTA Pure](https://primary.jwwb.nl/public/p/r/t/temp-clapbsszbkpnwzseqljn/20241120_184137-2-high.jpg?enable-io=true&enable=upscale&crop=1920%2C2570%2Cx0%2Cy0%2Csafe&width=245&height=328)
2. Prime the Pumps:
- Prime both A and B pumps using a syringe.
![Cytiva AKTA P9 pump](https://primary.jwwb.nl/public/p/r/t/temp-clapbsszbkpnwzseqljn/20241120_185725-high.jpg?enable-io=true&enable=upscale&crop=1920%2C1440%2Cx0%2Cy0%2Csafe&width=440&height=330)
3. Test the Pump Flow Restrictor Valves:
- Disconnect an outlet tube from the flow restrictor valve (.e.g. 2A) at the system pressure monitor.
- Observe for 1–2 minutes.
- Repeat the process for the other outlet, e.g. 2B.
- If a sample pump is installed, perform the test using the buffer inlet and disconnecting the sample pump flow restrictor valve (2S).
![AKTA Pure pressure monitor](https://primary.jwwb.nl/public/p/r/t/temp-clapbsszbkpnwzseqljn/image-high-985eii.png?enable-io=true&enable=upscale&crop=1440%2C1921%2Cx197%2Cy0%2Csafe&width=207&height=276)
4. Results:
- The presence of flow from either outlet while the pump is idle suggests a failure in the pump flow restrictor valve. Possible causes include:
- Blockage preventing valve closure.
- Defective seals.
Further investigation and potential valve replacement may be required.
Additional Notes:
- The procedure is demonstrated on an AKTA Pure system but is applicable to any AKTA Chromatography system with a pump flow restrictor valve installed. There are variations between system models such as a single outlet from the pump flow restrictor valve as opposed to two. It is important to familiarise yourself with the system flow path before attempting this procedure.
- Important: This test is specific to pump flow restrictor valves only. The system flow restrictor serves a different purpose and requires a separate procedure (refer to the relevant AKTA User Manual for details).
For further assistance, feel free to contact us.
Add comment
Comments