How to Perform a Siphoning Test for AKTA Chromatography Systems

Published on 10 January 2025 at 09:10
Cytiva AKTA Pure

What is Siphoning and Why is it Important?

Siphoning occurs in a chromatography system due to the tubing flow path's configuration. It happens when the outlet point is positioned lower than the reservoir, creating a vertical distance that drives flow under gravitational pressure. In chromatography systems, another scenario can occur where one pump is active while the other is idle. This can lead to a phenomenon where the active pump creates a vacuum or low-pressure area that causes fluid to be drawn from the inlet line of the idle pump.

To counteract siphoning:

  • Pump flow restrictor valves are installed.
  • These valves remain closed under gravitational pressure and only open when the pump creates enough pressure to overcome it.

Importance:

  • Prevents flow when pumps are idle.
  • Avoids unwanted contamination between the A & B inlet lines.

When to Perform a Siphoning Test

  • Annually: It is best practice to perform the test at least once a year.
  • When Suspected Issues Arise: Indications of a faulty pump flow restrictor valve include:
    • Increased intake from the reservoir.
    • Inconsistent purification results.
    • Unstable UV & Conductivity readings.

How to Perform the Siphoning Test

Performing the test is quick and straightforward, typically taking about 5 minutes. In addition to our step-by-step guide, we have created a short easy to follow video for your assistance.

Materials Required:

  • A container with DI water.
  • A suitably sized syringe to purge the system pumps.

1. Prepare the Setup:

  • Place a container filled with DI water on top of the instrument and Immerse inlets A1 and B1 into the container.
Cytiva AKTA Pure

2. Prime the Pumps:

  • Prime both A and B pumps using a syringe.
Cytiva AKTA P9 pump

3. Test the Pump Flow Restrictor Valves:

  • Disconnect an outlet tube from the flow restrictor valve (.e.g. 2A) at the system pressure monitor.
  • Observe for 1–2 minutes.
  • Repeat the process for the other outlet, e.g. 2B.
  • If a sample pump is installed, perform the test using the buffer inlet and disconnecting the sample pump flow restrictor valve (2S).
AKTA Pure pressure monitor

4. Results:

  • The presence of flow from either outlet while the pump is idle suggests a failure in the pump flow restrictor valve. Possible causes include:
    • Blockage preventing valve closure.
    • Defective seals.

Further investigation and potential valve replacement may be required.

Additional Notes:

  • The procedure is demonstrated on an AKTA Pure system but is applicable to any AKTA Chromatography system with a pump flow restrictor valve installed. There are variations between system models such as a single outlet from the pump flow restrictor valve as opposed to two. It is important to familiarise yourself with the system flow path before attempting this procedure.
  • Important: This test is specific to pump flow restrictor valves only. The system flow restrictor serves a different purpose and requires a separate procedure (refer to the relevant AKTA User Manual for details).

 

For further assistance, feel free to contact us.

Add comment

Comments

There are no comments yet.